Problem in Operation
1. If the meter does not display a message after inserting
a test strip:
2. If the test does not start after applying the sample:
PROBABLE CAUSE WH AT TO DO
Battery exhausted. Replace the battery .
Battery incorrectly installed or
absent .
Check that the battery is cor -
rectly installed.
Test strip inserted upside-
down or incompletely.
Insert the test strip correctly
and “ ” facing up.
Defective meter. Please call the dealer for
service.
PROBABLE CAUSE WH AT TO DO
Repeat the test using a new
test strip with larger volume
of blood sample.
Defective test strip. Repeat the test with a new
test strip .
Sample applied after
automatic shutoff
( 3 minutes after last user
action).
Repeat the test with a new
test strip. Apply sample only
when flashing
appears
on the display.
Defective meter
.
Please call the dealer for
service.